Stretch tests - 4mbar to-350mbar
Description
Experiments were done to determine the pressure-strain dependency of a silicone membrane in a custom made stretching device. 1 µm Ø Dragon Green fluorescent polystyrene microbeads (Bangs Laboratories, Inc., Fishers, IN, USA) were adsorbed to the surface of the silicone membrane. ZEISS Axio Scope microscope with A-Plan 10x/0.25 Ph 1 objective was used to take image sequences of membrane deformation. The membrane was deformed by decreasing the pressure from -4 mbar to -350 mbar 75 mbar steps. The microscope was manually refocused after each step. Within one strain measurement, three locations near the edge of the cell culture well (imaging locations locations 1-3) and one location in the center (imaging location4) were measured. Three to five repeated image sequences were taken at each location.For analysis, Particle Tracker 2D/3D plugin (Mosaic) [29] in Fiji [30] was used to track the trajectories of the beads (data not shown).
Show moreYear of publication
2022
Type of data
Project
Other information
Fields of science
Biochemistry, cell and molecular biology; Medical biotechnology
Language
English
Open access
Open